AIR-COOLED PROGRAMMABLE DC ELECTRONIC LOAD SPECIFICATIONS |
CV MODE SPECIFICATIONS |
|
CC MODE SPECIFICATIONS |
| CVL RANGE |
0 ~ 10% of Vmax |
CCL RANGE |
0 ~ 10% of Imax |
| CVM RANGE |
0 ~ 50% of Vmax |
CCM RANGE |
0 ~ 50% of Imax |
| CVH RANGE |
0 ~ 100% of Vmax |
CCH RANGE |
0 ~ 100% of Imax |
| ACCURACY |
0.05% of Value ± 0.1% of Rating |
ACCURACY |
0.05% of Value ± 0.1% of Rating |
| RESOLUTION |
1/16000 of Rated Voltage |
RESOLUTION |
1/16000 of Rated Voltage |
| TRANSIENT TIME (SLOW) |
0.500 ~511.9 (ms) |
TRANSIENT TIME (SLOW) |
0.500 ~511.9 (ms) |
| TRANSIENT TIME (FAST) |
0.500 ~51.19 (ms) |
TRANSIENT TIME (FAST) |
0.050 ~51.19 (ms) |
| |
|
|
|
CR and CP MODE SPECIFICATIONS |
PROTECTION |
| Please reference website datasheet for details |
OVER POWER PROTECTION |
110% * Pmax |
| |
|
OVER VOLTAGE PROTECTION |
105% * Vmax |
GENERAL SPECIFICATIONS |
OVER CURRENT PROTECTION |
110% * Imax |
| REMOTE INTERFACES |
RS-232, GPIB & USB/ETHERNET |
OVER TEMERPATURE PROTECTION |
90°C ± 5°C |
| ANALOG PROGRAMMING |
0 ~ 10Vdc |
REVERSE MAXIMUM CURRENT |
110% of Imax |
| ACCURACY |
Mode Accuracy ± 0.1% of Rating |
REMOTE INHIBIT (RI) |
Short |
| VMON ACCURACY |
0.10% of RDG ± 0.1% of Rating |
FAULT INDICATOR |
SPDT Relay |
| IMON ACCURACY |
0.10% of RDG ± 0.1% of Rating |
OTHER PROGRAMMABLE PROTECTIONS: OPP, OVP, OCP, UVL & ANTI-OSCILLATION |
| FREQUENCY RANGE |
0.1Hz ~ 10kHz |
|
|
| DUTY RANGE |
1.000 ~100.0% |
DIELECTRIC STRENGTH |
| FREQUENCY & DUTY ACCURACY |
0.1% of Setting |
PRIMARY CIRCUIT TO CHASSIS |
1500Vac for 1 Minute |
| AC INPUT |
95~240Vac 48 ~ 62Hz |
PRIMARY CIRCUIT TO LOAD TERMINAL |
1500Vac for 1 Minute |
| OPERATING TEMPERATURE |
5°C ~ 40°C |
LOAD TERMINAL TO CHASSIS |
1500Vdc for 1 Minute |