BENCH-TOP PROGRAMMABLE DC ELECTRONIC LOAD SPECIFICATIONS
CV MODE SPECIFICATIONS 
CC MODE SPECIFICATIONS 
CVL RANGE 0 ~ 10% of VMAX CCL RANGE 0 ~ 10% of IMAX
CVM RANGE 0 ~ 50% of VMAX CCM RANGE 0 ~ 50% of IMAX
CVH RANGE 0 ~ 100% of VMAX CCH RANGE 0 ~ 100% of IMAX
ACCURACY 0.05% of Value ± 0.1% of Rating  ACCURACY 0.05% of Value ± 0.1% of Rating 
RESOLUTION 1/16000 of Rated Voltage CCUL RANGE 0 ~ 10% of IMAX
TRANSIENT TIME (SLOW) 0.250 ~ 255.9 (ms) CCUL ACCURACY 0.05% of Value ± 0.05% of Range
TRANSIENT TIME (FAST) 0.250 ~ 25.59 (ms) RESOLUTION 1/16000 of Rated Current
    TRANSIENT TIME (SLOW) 0.250 ~ 255.9 (ms)
CR and CP MODE SPECIFICATIONS 
TRANSIENT TIME (FAST) 0.025 ~ 25.59 (ms)
Please reference website datasheet for details     
   
GENERAL SPECIFICATIONS 
PROTECTION 
REMOTE INTERFACES RS-232, GPIB & USB/ETHERNET OVER POWER PROTECTION 110% * PMAX
ANALOG PROGRAMMING 0 ~ 10Vdc OVER VOLTAGE PROTECTION 105% * VMAX
ACCURACY Mode Accuracy ± 0.1% of Rating  OVER CURRENT PROTECTION 110% * IMAX
VMON ACCURACY 0.10% of RDG ± 0.1% of Rating  OVER TEMERPATURE PROTECTION 90°C ± 5°C
IMON ACCURACY 0.10% of RDG ± 0.1% of Rating  REMOTE INHIBIT (RI) Short
IMON (CCUL) ACCURACY CCUL Mode Accuracy ± 0.1% of Rating  FAULT INDICATOR SPDT Relay
FREQUENCY RANGE 0.1Hz ~ 20kHz  Other Programmable Protections:  OPP, OVP, OCP, UVL & Anti-Oscillation
ACCURACY 0.10%    
AC INPUT 95~240Vac 48 ~ 62Hz
DIELECTRIC STRENGTH 
OPERATING TEMPERATURE 5°C ~ 40°C

PRIMARY CIRCUIT TO CHASSIS

1500Vac for 1 Minute
DIMENSIONS 15.65" (L) x 8.35" (W) x 5.20" (H)

PRIMARY CIRCUIT TO LOAD TERMINAL

1500Vac for 1 Minute
WEIGHT 22 lbs

LOAD TERMINAL TO CHASSIS

1500Vdc for 1 Minute
 
 
AMREL MANUFACTURES THE WIDEST RANGE OF ELECTRONIC TEST EQUIPMENT ON THE MARKET INCLUDING:
 
eLoad dc Electronic Loads
PLA Series of Air-cooled Programmable dc Electronic Loads
600w to 250kW
PLW Water-cooled Programmable dc Electronic Loads
3kW to 250kW+
Programmable dc 
Electronic Loads System Solutions
BPL Bench-top Programmable dc 
Electronic Loads
FEL/PEL Low-voltage Programmable dc Electronic Loads
60W to 300W
FEL/PEL Low-voltage Programmable dc Electronic Loads
60W to 600W
ZVL Zero-volt Programmable dc Electronic Loads
60W to 300W
ZVL Zero-volt Programmable dc Electronic Loads
ZVL Zero-volt Programmable dc Electronic Loads
200W to 1.5kW
ePower dc Power Supplies
SPD Dual-channel Programmable dc Switch Mode Power Supplies
300W to 360W

1.2kW
SPS Programmable dc Switch Mode Power Supplies
1.6kW

30kW to 900kW+
FRA Frequency Response Analyzer
100W to 200W
FRA Frequency Response Analyzer
15W to 2kW
MC-1 RS-485 Multi-channel Controller  
      MCU Ethernet Multi-channel Controller
up to 200W per Channel

Complete AMREL's request information form
Call toll free (800) 654-9838 U.S only or (626) 443-6818 Intl. Monday-Friday 8am-5pm (PST) to speak with an AMREL representative.
Visit AMREL’s ePower FAQs or eLoad FAQs for commonly asked product questions.
Copyright © 2009 AMREL All Rights Reserved 3445 Fletcher Avenue El Monte, CA 91731 Phone 626-443-6818 Fax 626-443-8600